IV. CLOSED LOOP CONTROL
In “Leaks: the Good, the Bad and the Ugly” we saw
how closed loop control is applied in a mass flow
controller with the control input being supplied by the
flow sensor with a solenoid proportioning valve as the
variable flow control element. The same approach can
be applied on a larger scale to a vacuum system. Figure
4 on the next page represents an extension to the
previous example with closed loop pressure control.
This particular configuration is referred to as upstream
pressure control. This is because we are driving a valve
that is upstream of the chamber.
Proportioning valves are an important element in
vacuum systems. Figure 5 depicts the general
configuration of a solenoid proportioning valve.
This type of valve uses the magnetic field
developed by a solenoid to move a magnetic armature
the Bell Jar, Vol. 10, No. 3/4, Summer/Fall 2001
19
Figure 4 -
Figure 2 - Ace Glass 8741 Vacuum Regulator. “Increased sensitivity
and versatility over other models, mercury requirement is reduced to
20-30cc. A control point within 1mm Hg can be fine-tuned by adjusting
the height of the pump-out tubulation. The Cartesian driver float drops
away from pump-out orifice on an increas- ing system pressure and
reseals it when the reference pressure set within large bulb is
reattained. Reference pressure is easily set. The spherical joint allows
the attitude of the orifice to be adjusted to accomodate a small
continuous leak without hunting of the float.” $430.12 in 2007.
Vacuum Chamber
Pressure
Gauge
Leak
Valve
Throttle Valve
To Pump
Gas
Supply
Figure 3 - Manual Pressure Control Elements